Impact Hammer Calibration System
It consists of a standard impact accelerometer, a pendulum, a pendulum clamping and releasing mechanism, a standard mass block, a mass block suspension mechanism, a dynamic data collector and a testing software. Through  the  collision  between  the  pendulum  and  the  standard  mass  block,  the  outputs  of  the  reference accelerometer and force sensor are tested, and the conversion of acceleration and force is realised through F=MA to calibrate the dynamic force sensor sensitivity, linearity and other indexes. By adjusting the fall height of the pendulum, different ranges of calibration and linearity testing can be achieved.
Pulse width: 0.1ms-50ms (adjustable); Mass block: about 1kg; Impact acceleration:  ≥1000g; Impact force:  ≥10kN.